top of page

BCA-Plus

Digital Mobile C-Arm

BMI BCA-Plus

PRODUCT Specifications

Electrical Charachteristics

· Voltage Standard: 230 Vac (±10% ) single phase
Optional: 115 Vac (±10% ) single phase
· Frequency: 50/60 Hz
· Max. Absorbed current: 10 A @ 230 Vac in fluoro mode
32 A @ 230 Vac in radiography mode
· Line compensation: Automatic
· Line resistance: < 0.4 Ω (supply at 230 Vac)
· Standard mains plug: 16 A (for supply at 230 Vac)

Radiological DATA

Generator nominal power: (IEC 60601-2-7) 4 kW @ 100kV 5 kW @ 100kV
High frequency generator: 40 kHz 40 kHz
Max. ripple: < 2% < 2%
High voltage range (Rad & Fluoro): 40 ÷ 110 kV 40 ÷ 120 kV
Current range in Continuous Fluoroscopy: 0.2 ÷ 4 mA 0.2 ÷ 4 mA
Current range in Radiography: 25mA ÷ 80 mA 25 mA ÷ 100 mA
mAs range in Radiography: 0.5 ÷ 240 mAs 0.5 ÷ 300 mAs
Exposure time in Radiography: 0.02 ÷ 3 sec 0.02 ÷ 3 sec
Total filtration: > 3,2 mmAl > 3,2 mmAl

X-RAY Tube

Model F112 X20P
Anode Stationary Rotating
Anode slope: 12° 10°
Focal spot size according to IEC 336: 0.5 Small Focus / 1.5 Large Focus 0.3 Small Focus / 0.6 Large Focus
Anode power rating: 0.8 kW SF / 4.2 kW LF 5kW SF / 17kW LF
Max. Anode heat content: 30 kJ (40 kHU) 150 kJ (200 kHU)
Max. continuous anode thermal dissipation: 270 W (360 HU/s) 300 W (400 HU/s)

Monobloc

Monobloc thermal capacity: 1000 kJ (1340 kHU) 1000 kJ (1340 kHU)
Max. continuous thermal dissipation: 100 W, 133 HU/sec, 8000 HU/min 100 W, 133 HU/sec, 8000 HU/min

Image intensifier

Size: 9” or 12”
Model: E5830SD-P4A or E5796SD-P2A
Number of fields: 3 (9” / 6” / 4.5”) or 3 (12” / 9” / 6”)
Input nominal diameter: 230 mm or 310 mm
Central Resolution (typ.): 52 / 58 / 68 lp/cm  or46 / 50 / 56 lp/cm
Conversion factor: (cd*s/m2*μGy) 29 or 34
Contrast ratio: (10% area) 30:1 or 30:1
Contrast ratio: (10 mm dia.) 19:1 or 18:1
DQE (IEC Standard): 65% or 65%

Collimator

Standard type · Square (Radiography) and Circular (Fluoroscopy) fixed sizes,
and automatically interchangeable
Iris and parallel shutters type (Optional)
· Iris shutters: beam collimation from the smallest field to the selected I.I.
Imput field
· Semi transparent shutters: n°2 parallel shutters, Cu 0.6 mm, 360° rotating
Preview collimator (Optional)
Adjustment of the collimation on the displayed image without making x-ray.
Automatic positioning of the iris and parallel shutters to the selected position,
when x-ray is ON

Anti-Scatter GRID

Ratio 14:1 14:1
Focus 100 cm 100 cm
# of lines 60 lp/cm 60 lp/cm

CCD Camera 1k x 1k

Matrix 1024 x 1024 pixels
Technology interline transfer CCD progressive scan with microlenses
Acquisition 1024 x 1024 x 14 bits
Contrast resolution 16384 grey levels
Signal - Noise Ratio 61 dB

Digital image system

ACQUISITION

Image resolution 1024 x 1024 x 14 bits
Grey Levels 16384 grey levels
Number of images 1 + 3000
Number of images on Hard Disk 3000
Frame rate acquisition 15 fps

REAL time processing

Window, Level: YES
Contrast & Brightness: YES
Recursive Filters Noise reduction: 4 LEVELS (F0,F1,F2,F3)
Continuous Digital Rotation: YES
Horizontal and Vertical Image inversion: YES

POSt processing

Edge enhancement: YES
Grey scale inversion
(negative / positive): YES
Patients Directory: YES
Editing data patient: YES
Zoom (2x, 4x, 8x, 16x): YES
Overview windowing: Yes, 7 images with scrolling function
Image Snap-Shot: YES
USB and LAN socket: YES
Angle and distance Measurement: YES - Optional
Overlay texting and notes: YES - Optional
Preview Collimator: YES - Optional

UTILITY

Exporting SW for 1Kx1K images recording
(BMP for AVI format) on USB flash drive:
YES - Optional
Exporting SW for 1Kx1K images recording
(BMP for AVI format) on CD/DVD drive:
YES - Optional
Exporting SW for 1Kx1K images recording
(DICOM format) on USB flash drive:
YES - Optional
Exporting SW for 1Kx1K images recording
(DICOM format) on CD/DVD drive:
YES - Optional

DICOM Services

DICOM Store YES - Optional
DICOM Modality Worklist YES - Optional
DICOM Print YES - Optional
DICOM Storage Commitment YES - Optional
DICOM Query Retrieve YES - Optional
DICOM Modality Performed Procedure Step
(MPPS) YES - Optional
DICOM Radiation Dose Structured Report
(RDSR) YES - Optional

monitor

Size: 19” (483 mm)
Active area: 14.80” x 11.85” (376 x 301 mm)
Type TFT active matrix colour LCD, medical

CONNECTIVITY AND CONTROL

Analog video input VGA (DB15)
Digital video input DVI-I

ELECTRICAL CHARACTERISTICS

Standard power supply 110-230V
Line frequency 50-60 Hz

DISPLAY CHARACTERISTICS

Resolution: 1M Pixel (1280 x 1024)
Contrast: 800:1
Luminance: 270 cd/m2
Visibility angle: 178° H / 178° V
Pixel pitch: 0.294 mm
Color depth: 16.8 M
Dicom calibrated luminance: 250 cd/m2
Aspect ratio: 5:4
Display LUT: 10 bit
Display function Gamma, DICOM
ALC (Ambient light compensation): YES (from OSD - On Screen Display)
BLOS (Backlight Output Stabilization): YES

MECHANICAL CHARACTERISTICS

Dimensions: 430 x 384 x 85 mm
Weight: 9.8 kg
Mounting: VESA 100

APPROVALS

CEI 60601-1 12/98, CEI 60601-1 04/98, CEI 60601-1 02/98

Continuous fluoroscopy

Dose control (kV - mA): Automatic or manual
mA fluoro variation range: 0.2 ÷ 4 mA on Small Focus
kV - mA correlation: Automatic
Safety timer: Acoustic alarm resettable after 4’30” fluoroscopy exposure.
Automatic stop after 5 minutes of continuous X-ray emission

BOOST fluoroscopy

mA on all kV range: 8 mA fixed on Small Focus 12 mA fixed on Small Focus
X-ray pulse time: 30 msec
X-ray pulse rate: 6 Selections. 0.5 - 1 - 2 - 4 - 7.5 - 15 pulses/sec
Safety timer: Acoustic alarm resettable after 4’30” fluoroscopy exposure.
Automatic stop: after 5 minutes of continuous X-ray emission

HCFLD (High Contrast fluoro low dose)

mA: 6 mA fixed on Small Focus 8 mA fixed on Small Focus
X-ray pulse time: 30 msec
X-ray pulse rate: 6 Selections. 0.5 - 1 - 2 - 4 - 7.5 - 15 pulses/sec
Safety timer: 
Acoustic alarm resettable after 4’30” fluoroscopy exposure.
Automatic stop: after 5 minutes of continuous X-ray emission

Digital Radiography
(One Shot)

mA: 8 mA fixed on Small Focus

12 mA fixed on Small Focus
X-ray flash time 66 msec

Radiography

mA Range: 25mA ÷ 100 mA 25mA ÷ 100 mA
Max. current: 100 mA 100 mA
Exposure time: 0.02 ÷ 3 sec
Selection x-ray parameters: kV, mA, time

bottom of page